Bioreactance and fourth-generation pulse contour methods in monitoring cardiac index during off-pump coronary artery bypass surgery
Ylikauma, Laura Anneli; Ohtonen, Pasi Petteri; Erkinaro, Tiina Maria; Vakkala, Merja Annika; Liisanantti, Janne Henrik; Satta, Jari Uolevi; Juvonen, Tatu Sakari; Kaakinen, Timo Ilari (2021-05-26)
Ylikauma, L.A., Ohtonen, P.P., Erkinaro, T.M. et al. Bioreactance and fourth-generation pulse contour methods in monitoring cardiac index during off-pump coronary artery bypass surgery. J Clin Monit Comput 36, 879–888 (2022). https://doi.org/10.1007/s10877-021-00721-0
© The Author(s) 2021. This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
https://creativecommons.org/licenses/by/4.0/
https://urn.fi/URN:NBN:fi-fe2022092059588
Tiivistelmä
Abstract
The pulmonary artery catheter (PAC) is considered the gold standard for cardiac index monitoring. Recently new and less invasive methods to assess cardiac performance have been developed. The aim of our study was to assess the reliability of a non-invasive monitor utilizing bioreactance (Starling SV) and a non-calibrated mini-invasive pulse contour device (FloTrac/EV1000, fourth-generation software) compared to bolus thermodilution technique with PAC (TDCO) during off-pump coronary artery bypass surgery (OPCAB). In this prospective study, 579 simultaneous intra- and postoperative cardiac index measurements obtained with Starling SV, FloTrac/EV1000 and TDCO were compared in 20 patients undergoing OPCAB. The agreement of data was investigated by Bland–Altman plots, while trending ability was assessed by four-quadrant plots with error grids. In comparison with TDCO, Starling SV was associated with a bias of 0.13 L min−1 m−2 (95% confidence interval, 95% CI, 0.07 to 0.18), wide limits of agreement (LOA, − 1.23 to 1.51 L min−1 m−2), a percentage error (PE) of 60.7%, and poor trending ability. In comparison with TDCO, FloTrac was associated with a bias of 0.01 L min−1 m−2 (95% CI − 0.05 to 0.06), wide LOA (− 1.27 to 1.29 L min−1 m−2), a PE of 56.8% and poor trending ability. Both Starling SV and fourth-generation FloTrac showed acceptable mean bias but imprecision due to wide LOA and high PE, and poor trending ability. These findings indicate limited reliability in monitoring cardiac index in patients undergoing OPCAB.